Belfort Instrument
The ES-642 Remote Dust Monitor

Outdoor Particulate · Particulate Sensors

The ES-642 Remote Dust Monitor

The ES-642 Remote Dust Monitor from Belfort Instrument. Rugged Weatherproof Enclosure: For reliable continuous outdoor operation. Purge Air System: Keeps optical components clean so the ES-642 can operate in adverse environments without performance degradation.. Automatic Zero Calibration: Reduces long term measurement drift by automatically adjusting the zero value each hour. The interval is user settable from 1 to 999 minutes.

Features

  • Rugged Weatherproof Enclosure: For reliable continuous outdoor operation
  • Purge Air System: Keeps optical components clean so the ES-642 can operate in adverse environments without performance degradation.
  • Automatic Zero Calibration: Reduces long term measurement drift by automatically adjusting the zero value each hour. The interval is user settable from 1 to 999 minutes.
  • Controlled Inlet Heater: Active inlet humidity control to reduce measurement errors. The ES-642 monitors inlet humidity levels and regulates heater power to maintain lower humidity levels.
  • Front Panel LCD Display: Provides information about particulate concentration, flow rate and instrument status.
  • Flexible Mounting Hardware: Mounting bracket included for wall or 2” vertical mast mounting. Optional hardware available for vertical mast diameters up to 4.7 inches (120 mm).
  • Building Automation
  • Military Applications
  • Environmental Clean-Up Sites
  • Air Pollution Level Monitoring
  • Dust Level Warning Systems

Product Specifications

Measurement Principle:Particulate concentration by forward light scatter laser Nephelometer
Available Cut Points:TSP Inlet Standard. PM10, PM2.5, and PM1 sharp-cut cyclone inlets available.
Measurement Range:0 to 100 mg/m3 (0 to 100,000 mg/m3)
Measurement Sensitivity:.001 mg/m3.
Nephelometer Accuracy:± 5% traceable standard with 0.6um PSL.
Particle Size Sensitivity:0.1 to 100 micron.
Display:2 X 16 backlit LCD. Provides concentration, flow and status information.
Zero Calibration:Automatic Zero Calibration every hour or as programmed from 1 to 999 minutes.
Flow Rate:2.0 liters/minute ± 0.1 lpm.
Input Voltage:11 – 40 VDC
Power Consumption:12 VDC = 300 mA (heater off), 1.2 A (heater on) 24 VDC = 140 mA (heater off), 0.6 A (heater on)
Analog Output:4-20 mA and 0 – 10 VDC.
Serial Ports:RS-232 and RS-485
Serial Protocols:ASCII Text data format and MODBUS RTU.
Alarm Output:Normally open and normally closed relay 30 VDC @ 1A maximum.
Operating Temperature:-10 to +50°C. (Ambient Temperature Sensor Range -30 to +50°C).
Barometric Pressure:600 to 1040 mbar pressure sensor range.
Ambient Humidity Range:0 to 90% RH, non-condensing.
Intake Moisture Control:Automatic 10 Watt inlet heater module controlled to sample RH set point.
Factory Serivce Interval:24 Months typical, under continuous use in normal ambient air.
Mounting Options:Mounting bracket standard (wall or 2” vertical mast). Optional EX-905 tripod.
Unit Weight:2.7 kg (6.0 lbs.)
Unit Dimensions:22.9cm high, 17.8cm wide, 10.8cm deep. (9.0” x 7.0” x 4.25”). w/out inlet assy. 48.3cm high, 17.8cm wide, 10.8cm deep. (19.0” x 7.0” x 4.25”). w/ inlet assy.

Specifications are subject to change. Contact Belfort Instrument to confirm current specifications.

Need pricing or a recommendation?

Our engineers respond to quote requests quickly. Call (410) 342-2626.

Request a Quote