Ambient Meteorological · Pressure
092 Barometric Pressure Sensor
The Model 092 Barometric Pressure Sensor measures absolute atmospheric pressure and converts it into a linear, proportional voltage, using digital computer technology.
Features
The 092 features robust, low maintenance construction that will withstand a wide-range of adverse weather conditions, and consistently collects reliable barometric pressure data.
- Compact size
- Weatherproof enclosure
- Digital and analog outputs
- Permanent calibration; no service required
- Customer configured output
Operation
The enclosure of the 092 sensor houses a solid-state pressure transducer, with linearization and amplification instruments, and a low power embedded CPU.
The standard range of the 092 is 600 to 1100 hPa, which makes it suitable for elevations sea level to 10,000 ft. Other ranges can be configured with DIP switches on the board of the barometer. The digital range is always 600 to 1100 hPa.e.
Construction
The 092 is contained within a small polycarbonate enclosure with metal weather shield. It may be mounted outside or inside a building.
Small inlet holes allow the atmospheric pressure access to the sensing element.
Product Specifications
| PERFORMANCE CHARACTERISTICS | ||
|---|---|---|
| Range | 600 to 1100 hPa (17.72 to 32.48 inch/hg) | |
| Elevation | Sea level to 10,000 ft. (3048 m) | |
| Resolution | 0.1 hPa | |
| Temp. Operating Range | -40°C to 55°C (-40°F to 131°F) | |
| Temp. Compensated Range | -40°C to 55°C (-40°F to 131°F) | |
| Accuracy | ±0.35 hPa @ 20°C (68°F) ±1.0 hPa (±0.03 in Hg) over full range or ±0.5 hPa over any 200 hPa range | |
| Long Term Stability | ±1.0 hPa in one year | |
| ELECTRICAL CHARACTERISTICS | ||
| Analog Output | 0-1, 0-2, 0-2.5 or 0-5 VDC (Analog output automatically adjusts from zero to full scale for range selected | |
| Digital Output | RS-232, RS-485 & SDI-12 | |
| Digital Protocol | ASCII Terminal Mode RTU for RS-232 and RS-485 | |
| Baud Rates | 1200, 2400, 4800, 9600, & 19.2K | |
| Power Requirement | 10 mA @ 12 VDC, Typical | |
| Power Range | 6-16 VDC | |
| PHYSICAL CHARACTERISTICS | ||
| Weight | 8.8 oz. (250 g) | |
| Dimensions | 4.72 x 3.14 x 2.16 in (120 x 80 x 55 mm) | |
| CABLE | ||
| PN 1169 | Cable Assembly; specify length in feet or meters | |
Specifications are subject to change. Contact Belfort Instrument to confirm current specifications.